Inductively Coupled Plasma Emission Spectrometer AE-ICP700 | AELAB

The AELAB AE-ICP700 Inductively Coupled Plasma Emission Spectrometer is a laboratory instrument for elemental analysis by atomic emission.

Description

Product Overview

The AELAB AE-ICP700 Inductively Coupled Plasma Emission Spectrometer is a laboratory instrument for elemental analysis by atomic emission.

Applications

  • Elemental analysis of metals, alloys, and aqueous samples
  • Quantitative determination of trace elements
  • Quality control in industrial manufacturing
  • Research and development in materials science
  • Environmental monitoring of water and soil
  • Chemical and petrochemical analysis

Technical Specifications

Product Specifications and Technical Indicators

ParameterValue
1. Power Supply System
Working frequency27.12MHz
Frequency stability<0.05%
Matching methodAutomatic matching
Output power800W ~1600W, continuously adjustable, power supply efficiency greater than 65%
Output power stability≤0.05%
Output working coil inner diameter25mm, 3 turns, equipped with three-concentric quartz torch tube with outer diameter 20mm; coaxial nebulizer outer diameter 6mm; double-cylinder spray chamber outer diameter 35mm
2. Spectroscopic System
Optical pathCzerny turner type
Focal length1000mm
Grating specificationsIon-etched holographic grating, line density 3600L/mm or 2400L/mm; ruled area (80×110) mm

FAQ

What is the working frequency of the AE-ICP700?

The working frequency is 27.12 MHz.

What type of optical system does it use?

It uses a Czerny turner type optical path with 1000 mm focal length.

What grating specifications are available?

An ion-etched holographic grating with line density 3600 L/mm or 2400 L/mm; ruled area (80×110) mm.

Is the output power adjustable?

Yes, output power is continuously adjustable from 800 W to 1600 W with power stability ≤0.05%.

What torch and nebulizer are included?

A three-concentric quartz torch tube (outer diameter 20 mm), coaxial nebulizer (outer diameter 6 mm), and double-cylinder spray chamber (outer diameter 35 mm) are included.

What is the frequency stability?

The frequency stability is <0.05%.

Additional information

Model

AE-ICP700

Reviews

There are no reviews yet.

Be the first to review “Inductively Coupled Plasma Emission Spectrometer AE-ICP700 | AELAB”

Your email address will not be published. Required fields are marked *